Performance of a Chamber for Studying the Liquid Xenon Response to Nuclear Recoils

dc.creatorChepel, V.
dc.creatorNeves, F.
dc.creatorSolovov, V.
dc.creatorPereira, A.
dc.creatorLopes, M. I.
dc.creatorda Cunha, J. Pinto
dc.creatorMendes, P.
dc.creatorLindote, A.
dc.creatorSilva, C. P.
dc.creatorMarques, R. Ferreira
dc.creatorPolicarpo, A. J. P. L.
dc.date2005-01-07
dc.date2005-01-14
dc.date.accessioned2026-07-07T12:17:03Z
dc.date.available2026-07-07T12:17:03Z
dc.descriptionThe design and performance of a 1.2 liter liquid xenon chamber equipped with 7 two-inch photomultiplier tubes, with the purpose of studying the scintillation response of xenon to gamma-rays and neutrons, is described. Measurements with gamma-rays indicate a high VUV light collection efficiency resulting in ~5.5 photoelectrons per 1 keV of deposited energy. The energy resolution (FWHM) is 18% and 22%, for 122 keV and 511 keV gamma-rays, respectively. An algorithm for the reconstruction of the scintillation coordinates in (x,y) plane was developed and tested. The position resolution is estimated to be 6.9 mm (sigma) for 122 keV gamma-rays.
dc.description6 pages, 6 figures; Presented at IEEE Nuclear Science Symposium and Medical Imaging Conference, Rome 2004; Submitted to IEEE Transactions on Nuclear Science
dc.identifierhttps://arxiv.org/abs/physics/0501036
dc.identifierhttp://arxiv.org/abs/physics/0501036
dc.identifierIEEETrans.Nucl.Sci.52:2793-2800,2005
dc.identifierdoi:10.1109/TNS.2005.862782
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/211997
dc.subjectInstrumentation and Detectors
dc.subjectAstrophysics
dc.subjectSpace Physics
dc.titlePerformance of a Chamber for Studying the Liquid Xenon Response to Nuclear Recoils
dc.typetext

Files

Collections