A stiffness sensor to help in the diagnosis and the surgery of orbital pathologies

dc.creatorLuboz, Vincent
dc.creatorAmbard, Dominique
dc.creatorBoutault, Franck
dc.creatorSwider, Pascal
dc.creatorPayan, Yohan
dc.date2006-10-22
dc.date.accessioned2026-07-07T07:29:58Z
dc.date.available2026-07-07T07:29:58Z
dc.descriptionProptosis is characterized by a protrusion of the eyeball due to an increase of the orbital tissue volume. To recover a normal eyeball positioning, the most frequent surgical technique (BROD technique) consists in the osteotomy of orbital walls combined with a loading on the eyeball to initiate tissue decompression. In this paper, a stiffness sensor device is proposed to (1) provide to the surgeon pre, intra and post-operative data concerning the stiffness of the intra-orbital soft tissues, and (2) provide constitutive parameters to the Finite Element model of the intra-orbital tissues already developed by the authors and used to predict consequences orbital surgery.
dc.identifierhttps://arxiv.org/abs/physics/0610174
dc.identifierhttp://arxiv.org/abs/physics/0610174
dc.identifierProceedings of the Second International Conference Surgetica'2005: Computer-Aided Medical Interventions: tools and applications. Sauramps Medical Editors (Ed.) (2005) 147-154
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/118316
dc.subjectMedical Physics
dc.titleA stiffness sensor to help in the diagnosis and the surgery of orbital pathologies
dc.typetext

Files

Collections