Development of a 5 MHz frequency difference pre-multiplier for a short term frequency stability bench of the oscillators

dc.creatorSalzenstein, Patrice
dc.creatorJouvenceau, Xavier
dc.creatorVacheret, Xavier
dc.creatorMartin, Gilles
dc.creatorLardet-Vieudrin, Franck
dc.date2007-06-13
dc.date.accessioned2026-07-07T08:09:48Z
dc.date.available2026-07-07T08:09:48Z
dc.descriptionThis paper reports the realization of a 5 MHz frequency difference x10 pre-multiplier, developed in the laboratory to replace an obsolete one. The principle we chose is to synthesize a 45 MHz and a 50 MHz from a reference signal and from the signal to be measured, and to subtract one to the other to generate a 5 MHz, whose precision on the measurement is increased by a factor 10. Obtained Allan variance y() at 1 s is 5.10-14 and output spectral density of phase noise floor is about -160 dBc/Hz at 5 MHz.
dc.identifierhttps://arxiv.org/abs/0706.1853
dc.identifierhttp://arxiv.org/abs/0706.1853
dc.identifierProceedings of the IEEE Frequency Control Symposium & European Frequency and Time Forum (01/06/2007) NA
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/131645
dc.subjectInstrumentation and Detectors
dc.titleDevelopment of a 5 MHz frequency difference pre-multiplier for a short term frequency stability bench of the oscillators
dc.typetext

Files