Development of a 5 MHz frequency difference pre-multiplier for a short term frequency stability bench of the oscillators
| dc.creator | Salzenstein, Patrice | |
| dc.creator | Jouvenceau, Xavier | |
| dc.creator | Vacheret, Xavier | |
| dc.creator | Martin, Gilles | |
| dc.creator | Lardet-Vieudrin, Franck | |
| dc.date | 2007-06-13 | |
| dc.date.accessioned | 2026-07-07T08:09:48Z | |
| dc.date.available | 2026-07-07T08:09:48Z | |
| dc.description | This paper reports the realization of a 5 MHz frequency difference x10 pre-multiplier, developed in the laboratory to replace an obsolete one. The principle we chose is to synthesize a 45 MHz and a 50 MHz from a reference signal and from the signal to be measured, and to subtract one to the other to generate a 5 MHz, whose precision on the measurement is increased by a factor 10. Obtained Allan variance y() at 1 s is 5.10-14 and output spectral density of phase noise floor is about -160 dBc/Hz at 5 MHz. | |
| dc.identifier | https://arxiv.org/abs/0706.1853 | |
| dc.identifier | http://arxiv.org/abs/0706.1853 | |
| dc.identifier | Proceedings of the IEEE Frequency Control Symposium & European Frequency and Time Forum (01/06/2007) NA | |
| dc.identifier.uri | http://salesiana.dossiersoluciones.com/handle/123456789/131645 | |
| dc.subject | Instrumentation and Detectors | |
| dc.title | Development of a 5 MHz frequency difference pre-multiplier for a short term frequency stability bench of the oscillators | |
| dc.type | text |