Photosensitive Gaseous Detectors for Cryogenic Temperature Applications

dc.creatorPeriale, L.
dc.creatorPeskov, V.
dc.creatorIacobaeus, C.
dc.creatorLund-Jensen, B.
dc.creatorPicchi, P.
dc.creatorPietropaolo, F.
dc.creatorRodionov
dc.date2005-09-09
dc.date.accessioned2026-07-07T11:07:41Z
dc.date.available2026-07-07T11:07:41Z
dc.descriptionThere are several proposals and projects today for building LXe Time Projection Chambers (TPCs) for dark matter search. An important element of these TPCs are the photomultipliers operating either inside LXe or in vapors above the liquid. We have recently demonstrated that photosensitive gaseous detectors (wire type and hole-type) can operate perfectly well until temperatures of LN2. In this paper results of systematic studies of operation of the photosensitive version of these detectors (combined with reflective or semi-transparent CsI photocathodes) in the temperature interval of 300-150 K are presented. In particular, it was demonstrated that both sealed and flushed by a gas detectors could operate at a quite stable fashion in a year/time scale. Obtained results, in particular the long-term stability of photosensitive gaseous detectors, strongly indicate that they can be cheap and simple alternatives to photomultipliers or avalanche solid-state detectors in LXe TPC applications.
dc.descriptionSubmitted to the Proceedings of the PSD-7 Conf. in Liverpool, UK
dc.identifierhttps://arxiv.org/abs/physics/0509077
dc.identifierhttp://arxiv.org/abs/physics/0509077
dc.identifierNucl.Instrum.Meth.A573:302-305,2007
dc.identifierdoi:10.1016/j.nima.2006.10.264
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/189929
dc.subjectInstrumentation and Detectors
dc.titlePhotosensitive Gaseous Detectors for Cryogenic Temperature Applications
dc.typetext

Files