Bolometric calibration of a superfluid $^3$He detector for Dark Matter search: direct measurement of the scintillated energy fraction for neutron, electron and muon events

dc.creatorWinkelmann, C. B.
dc.creatorElbs, J.
dc.creatorBunkov, Yu. M.
dc.creatorCollin, E.
dc.creatorGodfrin, H.
dc.creatorKrusius, M.
dc.date2006-11-28
dc.date.accessioned2026-07-07T11:24:07Z
dc.date.available2026-07-07T11:24:07Z
dc.descriptionWe report on the calibration of a superfluid $^3$He bolometer developed for the search of non-baryonic Dark Matter. Precise thermometry is achieved by the direct measurement of thermal excitations using Vibrating Wire Resonators (VWRs). The heating pulses for calibration were produced by the direct quantum process of quasiparticle generation by other VWRs present. The bolometric calibration factor is analyzed as a function of temperature and excitation level of the sensing VWR. The calibration is compared to bolometric measurements of the nuclear neutron capture reaction and heat depositions by cosmic muons and low energy electrons. The comparison allows a quantitative estimation of the ultra-violet scintillation rate of irradiated helium, demonstrating the possibility of efficient electron recoil event rejection.
dc.description17 pages, submitted to Nuc. Instr. Meth. A
dc.identifierhttps://arxiv.org/abs/physics/0611273
dc.identifierhttp://arxiv.org/abs/physics/0611273
dc.identifierNucl.Instrum.Meth.A574:264-271,2007
dc.identifierdoi:10.1016/j.nima.2007.01.180
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/195125
dc.subjectInstrumentation and Detectors
dc.subjectAstrophysics
dc.subjectSuperconductivity
dc.titleBolometric calibration of a superfluid $^3$He detector for Dark Matter search: direct measurement of the scintillated energy fraction for neutron, electron and muon events
dc.typetext

Files