X-ray multi-energy radiography with scintillator-photodiode detectors
| dc.creator | Naydenov, S. V. | |
| dc.creator | Ryzhikov, V. D. | |
| dc.creator | Grinyov, B. V. | |
| dc.creator | Lisetskaya, E. K. | |
| dc.creator | Opolonin, A. D. | |
| dc.creator | Kozin, D. N. | |
| dc.date | 2002-06-05 | |
| dc.date | 2002-11-29 | |
| dc.date.accessioned | 2026-07-07T05:47:39Z | |
| dc.date.available | 2026-07-07T05:47:39Z | |
| dc.description | For reconstruction of the spatial structure of many-component objects, it is proposed to use multi-radiography with detection of X-ray by combined detector arrays using detectors of ``scintillator-photodiode'' type. A theoretical model has been developed of multi-energy radiography for thickness measurements of multi-layered systems and systems with defects. Experimental studies of the sensitivity, output signal of various inspection systems based on scintillators $ZnSe(Te)$ and $CsI(Tl)$, and object image reconstruction (with organics and non-ogranics materials singled out) have been carried out. | |
| dc.description | 6 pages, 7 figures, LaTeX2e | |
| dc.identifier | https://arxiv.org/abs/physics/0206014 | |
| dc.identifier | http://arxiv.org/abs/physics/0206014 | |
| dc.identifier.uri | http://salesiana.dossiersoluciones.com/handle/123456789/84737 | |
| dc.subject | Instrumentation and Detectors | |
| dc.subject | Medical Physics | |
| dc.subject | Physics and Society | |
| dc.title | X-ray multi-energy radiography with scintillator-photodiode detectors | |
| dc.type | text |