An Investigation on Cooling of CZT Co-Planar Grid Detectors
| dc.creator | Dawson, J. V. | |
| dc.creator | Montag, C. | |
| dc.creator | Reeve, C. | |
| dc.creator | Wilson, J. R. | |
| dc.creator | Zuber, K. | |
| dc.date | 2008-11-12 | |
| dc.date.accessioned | 2026-07-07T12:36:22Z | |
| dc.date.available | 2026-07-07T12:36:22Z | |
| dc.description | The effect of moderate cooling on CdZnTe semiconductor detectors has been studied for the COBRA experiment. Improvements in energy resolution and low energy threshold were observed and quantified as a function of temperature. Leakage currents are found to contribute typically $\sim$5 keV to the widths of photopeaks. | |
| dc.description | 14 pages, 9 figures. Accepted for publication in Nuclear Inst. and Methods in Physics Research, A | |
| dc.identifier | https://arxiv.org/abs/0811.1843 | |
| dc.identifier | http://arxiv.org/abs/0811.1843 | |
| dc.identifier | Nucl.Instrum.Meth.A599:209-214,2009 | |
| dc.identifier | doi:10.1016/j.nima.2008.11.013 | |
| dc.identifier.uri | http://salesiana.dossiersoluciones.com/handle/123456789/218068 | |
| dc.subject | Instrumentation and Detectors | |
| dc.title | An Investigation on Cooling of CZT Co-Planar Grid Detectors | |
| dc.type | text |