Quantitative 3w-Scanning Thermal Microscopy: Modelling the AC/DC coupling and the sample heat conduction

dc.creatorChapuis, P. -O.
dc.creatorSaha, S. Kumar
dc.creatorVolz, S.
dc.date2007-09-12
dc.date.accessioned2026-07-07T08:29:05Z
dc.date.available2026-07-07T08:29:05Z
dc.descriptionA way to increase the Scanning Thermal Microscope (SThM) sensitivity in the harmonic 3w mode is to heat the probe with an AC current sufficiently high to generate a coupling between the AC and the DC signals. We detail in this paper how to properly take into account this coupling with a Wollaston-probe SThM. We also show how to link correctly the thermal conductivity to the thermal conductance measured by the SThM.
dc.descriptionSubmitted on behalf of TIMA Editions (http://irevues.inist.fr/tima-editions)
dc.identifierhttps://arxiv.org/abs/0709.1835
dc.identifierhttp://arxiv.org/abs/0709.1835
dc.identifierDans Proceedings of 12th International Workshop on Thermal investigations of ICs - THERMINIC 2006, Nice : France (2006)
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/137844
dc.subjectMaterials Science
dc.titleQuantitative 3w-Scanning Thermal Microscopy: Modelling the AC/DC coupling and the sample heat conduction
dc.typetext