Baseband Detection of Bistatic Electron Spin Signals in Magnetic Resonance Force Microscopy (MRFM)
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Description
In single spin Magnetic Resonance Force Microscopy (MRFM), the objective is to detect the presence of an electron (or nuclear) spin in a sample volume by measuring spin-induced attonewton forces using a micromachined cantilever. In the OSCAR method of single spin MRFM, the spins are manipulated by an external rf field to produce small periodic deviations in the resonant frequency of the cantilever. These deviations can be detected by frequency demodulation followed by conventional amplitude or energy detection. In this paper, we present an alternative to these detection methods, based on optimal detection theory and Gibbs sampling. On the basis of simulations, we show that our detector outperforms the conventional amplitude and energy detectors for realistic MRFM operating conditions. For example, to achieve a 10% false alarm rate and an 80% correct detection rate our detector has an 8 dB SNR advantage as compared with the conventional amplitude or energy detectors. Furthermore, at these detection rates it comes within 4 dB of the omniscient matched-filter lower bound.
8 pages, 9 figures, revision of paper contains correction to a typo on the first page (introduction section)
8 pages, 9 figures, revision of paper contains correction to a typo on the first page (introduction section)