Performance of An Axial Gas Ionization Detector

dc.creatorAdhikari, S.
dc.creatorBasu, C.
dc.creatorSamanta, C.
dc.creatorBrahmachari, S. S.
dc.creatorDas, B. P.
dc.creatorBasu, P.
dc.date2006-10-11
dc.date.accessioned2026-07-07T10:26:18Z
dc.date.available2026-07-07T10:26:18Z
dc.descriptionAn axial gas ionization chamber has been fabricated for use as a $ΔE$ detector in heavy ion induced nuclear reactions. Different operating parameters such as gas type, pressure, anode voltage and anode structures have been optimized. The transparency of the anode structure is observed to play an important role in improving the energy resolution of the detector.
dc.description19 pages, 7 figures
dc.identifierhttps://arxiv.org/abs/nucl-ex/0610016
dc.identifierhttp://arxiv.org/abs/nucl-ex/0610016
dc.identifierIEEETrans.Nucl.Sci.53:2270,2006
dc.identifierdoi:10.1109/TNS.2006.878132
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/176834
dc.subjectNuclear Experiment
dc.titlePerformance of An Axial Gas Ionization Detector
dc.typetext