A Compact High-Energy Neutron Spectrometer
| dc.creator | Brooks, F. D. | |
| dc.creator | Buffler, A. | |
| dc.creator | Allie, M. S. | |
| dc.creator | Herbert, M. S. | |
| dc.creator | Nchodu, M. R. | |
| dc.creator | Jones, D. T. L. | |
| dc.creator | Smit, F. D. | |
| dc.creator | Nolte, R. | |
| dc.creator | Dangendorf, V. | |
| dc.date | 2007-03-23 | |
| dc.date.accessioned | 2026-07-07T07:53:26Z | |
| dc.date.available | 2026-07-07T07:53:26Z | |
| dc.description | A compact liquid organic neutron spectrometer (CLONS) based on a single NE213 liquid scintillator (5 cm diam. x 5 cm) is described. The spectrometer is designed to measure neutron fluence spectra over the energy range 2-200 MeV and is suitable for use in neutron fields having any type of time structure. Neutron fluence spectra are obtained from measurements of two-parameter distributions (counts versus pulse height and pulse shape) using the Bayesian unfolding code MAXED. Calibration and test measurements made using a pulsed neutron beam with a continuous energy spectrum are described and the application of the spectrometer to radiation dose measurements is discussed. | |
| dc.description | To be published in Radiation Protection Dosimetry | |
| dc.identifier | https://arxiv.org/abs/physics/0703211 | |
| dc.identifier | http://arxiv.org/abs/physics/0703211 | |
| dc.identifier.uri | http://salesiana.dossiersoluciones.com/handle/123456789/126254 | |
| dc.subject | Instrumentation and Detectors | |
| dc.title | A Compact High-Energy Neutron Spectrometer | |
| dc.type | text |