Electric and magnetic field optimization procedure for Penning trap mass spectrometers

dc.creatorBeck, D.
dc.creatorBlaum, K.
dc.creatorBollen, G.
dc.creatorDelahaye, P.
dc.creatorGeorge, S.
dc.creatorGuenaut, C.
dc.creatorHerfurth, F.
dc.creatorHerlert, A.
dc.creatorLunney, D.
dc.creatorSchweikhard, L.
dc.creatorYazidjian, C.
dc.date2008-05-29
dc.date.accessioned2026-07-07T12:30:06Z
dc.date.available2026-07-07T12:30:06Z
dc.descriptionSignificant systematic errors in high-precision Penning trap mass spectrometry can result from electric and magnetic field imperfections. An experimental procedure to minimize these uncertainties is presented for the on-line Penning trap mass spectrometer ISOLTRAP, located at ISOLDE/CERN. The deviations from the ideal magnetic and electric fields are probed by measuring the cyclotron frequency and the reduced cyclotron frequency, respectively, of stored ions as a function of the time between the ejection of ions from the preparation trap and their capture in the precision trap, which influences the energy of their axial motion. The correction parameters are adjusted to minimize the frequency shifts.
dc.description25 pages, 9 figures
dc.identifierhttps://arxiv.org/abs/0805.4549
dc.identifierhttp://arxiv.org/abs/0805.4549
dc.identifierNucl.Instrum.Meth.A598:635-641,2009
dc.identifierdoi:10.1016/j.nima.2008.09.019
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/216041
dc.subjectInstrumentation and Detectors
dc.subjectAtomic Physics
dc.titleElectric and magnetic field optimization procedure for Penning trap mass spectrometers