Fabrication and Performance Tests of a Segmented P-Type HPGe Detector
| dc.creator | King III, George S. | |
| dc.creator | Avignone III, Frank T. | |
| dc.creator | Cox, Christopher E. | |
| dc.creator | Hossbach, Todd W. | |
| dc.creator | Jennings, Wayne | |
| dc.creator | Reeves, James H. | |
| dc.date | 2007-05-31 | |
| dc.date.accessioned | 2026-07-07T11:56:53Z | |
| dc.date.available | 2026-07-07T11:56:53Z | |
| dc.description | A p-type HPGe detector has been segmented by cutting, with a diamond saw, and etching six circumferential grooves through the Li-diffused dead layer. The degree of segmentation was tested with the 88 keV gamma rays from a well-collimated source of 109Cd. The fraction of events, recognized as occurring in more than one segment, and rejected from the energy interval 2038 +/- 5 keV, was measured as 0.59. | |
| dc.description | Submitted to Nuclear Instruments and Methods, Section A | |