Gamma Ray Spectroscopy with Scintillation Light in Liquid Xenon

dc.creatorNi, K.
dc.creatorAprile, E.
dc.creatorGiboni, K. L.
dc.creatorMajewski, P.
dc.creatorYamashita, M.
dc.date2006-08-03
dc.date2006-08-04
dc.date.accessioned2026-07-07T10:42:57Z
dc.date.available2026-07-07T10:42:57Z
dc.descriptionScintillation light from gamma ray irradiation in liquid xenon is detected by two Hamamatsu R9288 photomultiplier tubes (PMTs) immersed in the liquid. UV light reflector material, PTFE, is used to optimize the light collection efficiency. The detector gives a high light yield of 6 photoelectron per keV (pe/keV), which allows efficient detection of the 122 keV gamma-ray line from Co-57, with a measured energy resolution of (8.8+/-0.6)% (sigma). The best achievable energy resolution, by removing the instrumental fluctuations, from liquid xenon scintillation light is estimated to be around 6-8% (sigma) for gamma-ray with energy between 662 keV and 122 keV.
dc.identifierhttps://arxiv.org/abs/physics/0608034
dc.identifierhttp://arxiv.org/abs/physics/0608034
dc.identifierJINST1:P09004,2006
dc.identifierdoi:10.1088/1748-0221/1/09/P09004
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/182132
dc.subjectInstrumentation and Detectors
dc.subjectAstrophysics
dc.subjectNuclear Experiment
dc.titleGamma Ray Spectroscopy with Scintillation Light in Liquid Xenon
dc.typetext