A Terraced Scanning Superconducting Quantum Interference Device Susceptometer with Sub-Micron Pickup Loops
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Description
Superconducting Quantum Interference Devices (SQUIDs) can have excellent spin sensitivity depending on their magnetic flux noise, pick-up loop diameter, and distance from the sample. We report a family of scanning SQUID susceptometers with terraced tips that position the pick-up loops 300 nm from the sample. The 600 nm - 2 um pickup loops, defined by focused ion beam, are integrated into a 12-layer optical lithography process allowing flux-locked feedback, in situ background subtraction and optimized flux noise. These features enable a sensitivity of ~70 electron spins per root Hertz at 4K.
See http://stanford.edu/group/moler/publications.html for an auxiliary document containing additional fabrication details and discussion
See http://stanford.edu/group/moler/publications.html for an auxiliary document containing additional fabrication details and discussion