Scintillation counter with MRS APD light readout
| dc.creator | Akindinov, A. | |
| dc.creator | Bondarenko, G. | |
| dc.creator | Golovin, V. | |
| dc.creator | Grigoriev, E. | |
| dc.creator | Grishuk, Yu. | |
| dc.creator | Mal'kevich, D. | |
| dc.creator | Martemiyanov, A. | |
| dc.creator | Ryabinin, M. | |
| dc.creator | Smirnitskiy, A. | |
| dc.creator | Voloshin, K. | |
| dc.date | 2006-03-27 | |
| dc.date.accessioned | 2026-07-07T07:07:48Z | |
| dc.date.available | 2026-07-07T07:07:48Z | |
| dc.description | START, a high-efficiency and low-noise scintillation detector for ionizing particles, was developed for the purpose of creating a high-granular system for triggering cosmic muons. Scintillation light in START is detected by MRS APDs (Avalanche Photo-Diodes with Metal-Resistance-Semiconductor structure), operated in the Geiger mode, which have 1 mm^2 sensitive areas. START is assembled from a 15 x 15 x 1 cm^3 scintillating plastic plate, two MRS APDs and two pieces of wavelength-shifting optical fiber stacked in circular coils inside the plastic. The front-end electronic card is mounted directly on the detector. Tests with START have confirmed its operational consistency, over 99% efficiency of MIP registration and good homogeneity. START demonstrates a low intrinsic noise of about 10^{-2} Hz. If these detectors are to be mass-produced, the cost of a mosaic array of STARTs is estimated at a moderate level of 2-3 kUSD/m^2. | |
| dc.description | 6 pages, 5 figures | |
| dc.identifier | https://arxiv.org/abs/physics/0603224 | |
| dc.identifier | http://arxiv.org/abs/physics/0603224 | |
| dc.identifier | Nucl.Instrum.Meth. A539 (2005) 172-176 | |
| dc.identifier | doi:10.1016/j.nima.2004.10.026 | |
| dc.identifier.uri | http://salesiana.dossiersoluciones.com/handle/123456789/110527 | |
| dc.subject | Instrumentation and Detectors | |
| dc.title | Scintillation counter with MRS APD light readout | |
| dc.type | text |