Demonstration of photomultiplier tube operation at 29 K

dc.creatorNikkel, James A.
dc.creatorLippincott, W. Hugh
dc.creatorMcKinsey, Daniel N.
dc.date2007-02-07
dc.date.accessioned2026-07-07T11:21:07Z
dc.date.available2026-07-07T11:21:07Z
dc.descriptionWe describe measurements of gain, dark current, and quantum efficiency obtained while cooling a Hamamatsu R5912-02-MOD photomultiplier tube from room temperature to 29 K. We found that the PMT operated normally down to 29 K, with a reduced gain and quantum efficiency at the lowest temperatures. Furthermore, we found that the dark count rate increased as the temperature decreased. We conclude that these PMTs appear to be adequate for the requirements of the CLEAN experiment.
dc.description3 pages, 5 figures
dc.identifierhttps://arxiv.org/abs/astro-ph/0702202
dc.identifierhttp://arxiv.org/abs/astro-ph/0702202
dc.identifierJINST2:P11004,2007
dc.identifierdoi:10.1088/1748-0221/2/11/P11004
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/194169
dc.subjectAstrophysics
dc.titleDemonstration of photomultiplier tube operation at 29 K
dc.typetext

Files