Test Stand Measurements for an ILC Polarimeter

dc.creatorKäfer, Daniela
dc.creatorEyser, Oleg
dc.creatorHelebrant, Christian
dc.creatorList, Jenny
dc.creatorVelte, Ulrich
dc.date2007-09-17
dc.date.accessioned2026-07-07T12:41:16Z
dc.date.available2026-07-07T12:41:16Z
dc.descriptionThe setup of two different small scale teststands for measurements regarding an electron Cherenkov detector as part of the ILC polarimeters is presented. Component measurements already carried out are analyzed and others, foreseen for the near future, are discussed. The larger one of the two teststands features the old Cherenkov detector of the SLD experiment, which will be used as a reference for a number of crucial measurements. Especially, the requirements for the non-linearity of the read-out chain are studied in greater detail and methods for its precise measurement before and during operation are being developed accordingly.
dc.description4 pages, 4 figures (incl. 5 sub-figures, in total: 9 .eps files) LCWS'07 proceedings
dc.identifierhttps://arxiv.org/abs/0709.2683
dc.identifierhttp://arxiv.org/abs/0709.2683
dc.identifierECONF C0705302:MDI04,2007
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/219712
dc.subjectInstrumentation and Detectors
dc.titleTest Stand Measurements for an ILC Polarimeter
dc.typetext

Files

Collections