Experimental setup for low-energy laser-based angle resolved photoemission spectroscopy

dc.creatorKoralek, J. D.
dc.creatorDouglas, J. F.
dc.creatorPlumb, N. C.
dc.creatorGriffith, J. D.
dc.creatorCundiff, S. T.
dc.creatorKapteyn, H. C.
dc.creatorMurnane, M. M.
dc.creatorDessau, D. S.
dc.date2007-06-07
dc.date.accessioned2026-07-07T08:04:32Z
dc.date.available2026-07-07T08:04:32Z
dc.descriptionA laser-based angle resolved photoemission (APRES) system utilizing 6 eV photons from the fourth harmonic of a mode-locked Ti:sapphire oscillator is described. This light source greatly increases the momentum resolution and photoelectron count rate, while reducing extrinsic background and surface sensitivity relative to higher energy light sources. In this review, the optical system is described, and special experimental considerations for low-energy ARPES are discussed. The calibration of the hemispherical electron analyzer for good low-energy angle-mode performance is also described. Finally, data from the heavily studied high T_c superconductor Bi2Sr2CaCu2O8+δ(Bi2212) is compared to the results from higher photon energies.
dc.descriptionPlease download final version from Journal-Ref
dc.identifierhttps://arxiv.org/abs/0706.1060
dc.identifierhttp://arxiv.org/abs/0706.1060
dc.identifierREVIEW OF SCIENTIFIC INSTRUMENTS 78, 053905 (2007)
dc.identifierdoi:10.1063/1.2722413
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/129994
dc.subjectOther Condensed Matter
dc.subjectMaterials Science
dc.subjectSuperconductivity
dc.titleExperimental setup for low-energy laser-based angle resolved photoemission spectroscopy
dc.typetext

Files

Collections