Characterization of the first true coaxial 18-fold segmented n-type prototype detector for the GERDA project
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Description
The first true coaxial 18-fold segmented n-type HPGe prototype detector produced by Canberra-France for the GERDA neutrinoless double beta-decay project was tested both at Canberra-France and at the Max-Planck-Institut fuer Physik in Munich. The main characteristics of the detector are given and measurements concerning detector properties are described. A novel method to establish contacts between the crystal and a Kapton cable is presented.
21 pages, 16 Figures, to be submitted to NIM A
21 pages, 16 Figures, to be submitted to NIM A