New frequency-modulation readout based on relaxation oscillations

dc.creatorFurlan, Miha
dc.creatorKirk, Eugenie
dc.creatorZehnder, Alex
dc.date2005-07-01
dc.date2005-12-16
dc.date.accessioned2026-07-07T06:43:21Z
dc.date.available2026-07-07T06:43:21Z
dc.descriptionScaling of multi-pixel cryogenic detectors for imaging becomes increasingly difficult with size due to complexity of readout circuitry and cryogenic constraints (thermal load from wiring). We propose and demonstrate a new readout scheme based on a highly stable RF oscillator composed of a superconducting tunnel junction which exhibits relaxation oscillations. The oscillation frequency is almost linear with the analog bias signal over a wide operation range. The frequency signals from different detectors can be combined into one single readout line. The current noise of an optimized circuit is about 5 pA/sqrt(Hz), which is comparable to standard SQUID amplifiers. We show experimental data from `stand-alone' operation as well as response to microcalorimeter X-ray signals.
dc.description3 pages, 2 figures
dc.identifierhttps://arxiv.org/abs/physics/0507005
dc.identifierhttp://arxiv.org/abs/physics/0507005
dc.identifierNucl. Instrum. Methods Phys. Res. A 559, 796 (2006)
dc.identifierdoi:10.1016/j.nima.2005.12.143
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/102382
dc.subjectInstrumentation and Detectors
dc.titleNew frequency-modulation readout based on relaxation oscillations
dc.typetext