A Simple Method for Rise-Time Discrimination of Slow Pulses from Charge-Sensitive Preamplifiers

dc.creatorTõke, Jan
dc.creatorQuinlan, Michael J.
dc.creatorGawlikowicz, Wojtek
dc.creatorSchröder, W. Udo
dc.date2008-04-29
dc.date.accessioned2026-07-07T11:50:23Z
dc.date.available2026-07-07T11:50:23Z
dc.descriptionPerformance of a simple method of particle identification via pulse rise time discrimination is demonstrated for slow pulses from charge-sensitive preamplifiers with rise times ranging from 10 ns to 500 ns. The method is based on a comparison of the amplitudes of two pulses, derived from each raw preamplifier pulse with two amplifiers with largely differing shaping times, using a fast peak-sensing ADC. For the injected charges corresponding to energy deposits in silicon detectors of a few tens of MeV, a rise time resolution of the order of 1 ns can be achieved. The identification method is applicable in particle experiments involving large-area silicon detectors, but is easily adaptable to other detectors with a response corresponding to significantly different pulse rise times for different particle species.
dc.description10 pages, 7 figures
dc.identifierhttps://arxiv.org/abs/0804.4696
dc.identifierhttp://arxiv.org/abs/0804.4696
dc.identifierNucl.Instrum.Meth.A595:460-463,2008
dc.identifierdoi:10.1016/j.nima.2008.07.024
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/203547
dc.subjectNuclear Experiment
dc.titleA Simple Method for Rise-Time Discrimination of Slow Pulses from Charge-Sensitive Preamplifiers
dc.typetext