Thin gap parallel mesh chamber: a sparkless high-rate detector
| dc.creator | Fonte, P. | |
| dc.creator | Ferreira-Marques, R. | |
| dc.creator | Ivaniouchenkov, Y. | |
| dc.creator | Peskov, V. | |
| dc.creator | Policarpo, A. | |
| dc.date | 1998-03-16 | |
| dc.date.accessioned | 2026-07-07T05:56:07Z | |
| dc.date.available | 2026-07-07T05:56:07Z | |
| dc.description | In this work we describe a parallel geometry gaseous detector constituted by a thin (0.6 mm) amplifying gap preceded by a drift region. The gap is delimited by a metal plate and by a wire mesh. The detector can count up to a rate of 10^7 Hz/mm2 at gain 1000 and retains at lower counting rates the high gains (close to 10^6) that are typical of parallel geometry detectors. Additionally it was found that in the thin amplifying gap the discharges seem to be self-quenched and that full sparks do not develop. This fact, together with the robustness of the electrode materials, makes this an extremely resilient high-rate detector. | |
| dc.description | 8 html pages, 10 gif figures. Submited to NIM A | |
| dc.identifier | https://arxiv.org/abs/physics/9803021 | |
| dc.identifier | http://arxiv.org/abs/physics/9803021 | |
| dc.identifier.uri | http://salesiana.dossiersoluciones.com/handle/123456789/87488 | |
| dc.subject | Instrumentation and Detectors | |
| dc.title | Thin gap parallel mesh chamber: a sparkless high-rate detector | |
| dc.type | text |