Performance of the triple-GEM detector with optimized 2-D readout in high intensity hadron beam

dc.creatorBondar, A.
dc.creatorBuzulutskov, A.
dc.creatorShekhtman, L.
dc.creatorSokolov, A.
dc.creatorVasiljev, A.
dc.date2001-10-19
dc.date.accessioned2026-07-07T05:46:23Z
dc.date.available2026-07-07T05:46:23Z
dc.descriptionMultiple-GEM detectors are considered to be good candidates for tracking devices in experiments with high hadronic background. We present the results of the triple-GEM detectors beam test in a high intensity pion beam. The detectors had an optimized two-dimensional readout with minimized strip capacitance. Such optimization permitted the starting point of the efficiency plateau down to a gain of 5000. The probability of GEM discharges induced by heavily ionizing particles has been measured as a function of gain: at a gain of 20000 it amounts to 10^(-11) per incident particle. Such a value will ensure safe operation of the detector in the conditions of forward region of the LHC experiments.
dc.descriptionzip-file, contains main file in LATEX and 9 figures in eps and epsi format
dc.identifierhttps://arxiv.org/abs/physics/0110054
dc.identifierhttp://arxiv.org/abs/physics/0110054
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/84318
dc.subjectInstrumentation and Detectors
dc.titlePerformance of the triple-GEM detector with optimized 2-D readout in high intensity hadron beam
dc.typetext