The use of Geant4 for simulations of a plastic beta-detector and its application to efficiency calibration
| dc.creator | Golovko, V. V. | |
| dc.creator | Iacob, V. E. | |
| dc.creator | Hardy, J. C. | |
| dc.date | 2008-06-05 | |
| dc.date | 2008-06-15 | |
| dc.date.accessioned | 2026-07-07T11:51:26Z | |
| dc.date.available | 2026-07-07T11:51:26Z | |
| dc.description | Precise beta-branching-ratio measurements are required in order to determine ft-values as a part of our program to test the Electroweak Standard Model via unitarity of the Cabibbo-Kobayashi-Moskawa matrix. For the measurements to be useful in this test, their precision must be close to 0.1 %. In a branching-ratio measurement, we position the radioactive sample between a thin plastic scintillator used to detect beta-particles, and a HPGe detector for gamma-rays. Both beta singles and beta-gamma coincidences are recorded. Although the branching ratio depends most strongly on the HPGe detector efficiency, it has some sensitivity to the energy dependence of the beta-detector efficiency. We report here on a study of our beta-detector response function, which used Monte Carlo calculations performed by the Geant4 toolkit. Results of the simulations are compared to measured beta-spectra from several standard beta-sources. | |
| dc.description | 9 pages, 8 figures; preprint submited to Nucl. Instr. and Meth. A, May 2008 | |
| dc.identifier | https://arxiv.org/abs/0806.0886 | |
| dc.identifier | http://arxiv.org/abs/0806.0886 | |
| dc.identifier | Nucl.Instrum.Meth.A594:266-272,2008 | |
| dc.identifier | doi:10.1016/j.nima.2008.06.025 | |
| dc.identifier.uri | http://salesiana.dossiersoluciones.com/handle/123456789/203898 | |
| dc.subject | Nuclear Experiment | |
| dc.title | The use of Geant4 for simulations of a plastic beta-detector and its application to efficiency calibration | |
| dc.type | text |