Characterization of a CCD array for Bragg spectroscopy

dc.creatorIndelicato, Paul
dc.creatorBigot, Eric-Olivier Le
dc.creatorTrassinelli, Martino
dc.creatorGotta, Detlev
dc.creatorHennebach, Maik
dc.creatorNelms, Nick
dc.creatorDavid, Christian
dc.creatorSimons, Leopold M.
dc.date2006-02-23
dc.date.accessioned2026-07-07T07:04:09Z
dc.date.available2026-07-07T07:04:09Z
dc.descriptionThe average pixel distance as well as the relative orientation of an array of 6 CCD detectors have been measured with accuracies of about 0.5 nm and 50 $μ$rad, respectively. Such a precision satisfies the needs of modern crystal spectroscopy experiments in the field of exotic atoms and highly charged ions. Two different measurements have been performed by illuminating masks in front of the detector array by remote sources of radiation. In one case, an aluminum mask was irradiated with X-rays and in a second attempt, a nanometric quartz wafer was illuminated by a light bulb. Both methods gave consistent results with a smaller error for the optical method. In addition, the thermal expansion of the CCD detectors was characterized between -105 C and -40 C.
dc.descriptionSubmitted to Review of Scientific Instruments
dc.identifierhttps://arxiv.org/abs/physics/0602159
dc.identifierhttp://arxiv.org/abs/physics/0602159
dc.identifierReview of Scientific Instruments 77 (04/2006) 043107
dc.identifierdoi:10.1063/1.2194485
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/109251
dc.subjectAtomic Physics
dc.titleCharacterization of a CCD array for Bragg spectroscopy
dc.typetext