2026-07-072026-07-07http://salesiana.dossiersoluciones.com/handle/123456789/86422We describe a device to conveniently measure the positions of the phase boundaries in a dilution refrigerator. We show how a simple modification of a standard capacitive level gauge (segmentation of one of the electrodes) permits a direct calibration of the capacitance versus phase boundary position. We compare this direct calibration with the indirect procedure that must be adopted for a conventional capacitive level gauge. The device facilitates the correct adjustment of the 3He/4He fraction in the dilution refrigerator.3 pagesInstrumentation and DetectorsPhase boundary detection for dilution refrigeratorstext