Abstract
Description
We present a photoacoustic resonant cell, the internal volume of which is ~ 0.5 cm^3. The cavity shape of the cell is so adapted in order to minimize the background signal arising due to absorption of laser beam in the cell windows. An experimental setup, measurement procedure and design of the cell are described. Results of detection of ammonia in nitrogen flow with the help of a CO2-laser are represented. The minimal detectable absorption for the cell is estimated in the experiment to be ~ 5.1 10^(-8) cm^(-1) W Hz^(-1/2)
10 pages, 3 figures, report to conference on Quantum Electronics (13-16 October, 2008, Minsk, Belarus)
10 pages, 3 figures, report to conference on Quantum Electronics (13-16 October, 2008, Minsk, Belarus)