Application of a miniaturized photoacoustic cell for high-sensitivity laser detection of ammonia in gas media

dc.creatorGorelik, A. V.
dc.creatorStarovoitov, V. S.
dc.date2008-10-02
dc.date2008-12-29
dc.date.accessioned2026-07-07T12:22:14Z
dc.date.available2026-07-07T12:22:14Z
dc.descriptionWe present a photoacoustic resonant cell, the internal volume of which is ~ 0.5 cm^3. The cavity shape of the cell is so adapted in order to minimize the background signal arising due to absorption of laser beam in the cell windows. An experimental setup, measurement procedure and design of the cell are described. Results of detection of ammonia in nitrogen flow with the help of a CO2-laser are represented. The minimal detectable absorption for the cell is estimated in the experiment to be ~ 5.1 10^(-8) cm^(-1) W Hz^(-1/2)
dc.description10 pages, 3 figures, report to conference on Quantum Electronics (13-16 October, 2008, Minsk, Belarus)
dc.identifierhttps://arxiv.org/abs/0810.0410
dc.identifierhttp://arxiv.org/abs/0810.0410
dc.identifier.urihttp://salesiana.dossiersoluciones.com/handle/123456789/213582
dc.subjectOptics
dc.subjectInstrumentation and Detectors
dc.titleApplication of a miniaturized photoacoustic cell for high-sensitivity laser detection of ammonia in gas media
dc.typetext